On the orientation independence of the HYPERSCINT scintillation dosimetry research platform in a MR-linac environment.

The purpose of this work was to characterize the HYPERSCINT scintillation dosimetry research platform in a MR-linac environment, particularly with respect to its orientation. This study shows that the HYPERSCINT scintillation dosimetry platform can be used regardless of its orientation in a magnetic field environment. Together with its linearity to dose and dose-rate, the detector shows great promises for development of dosimetry solutions dedicated to the MR-Linac environment.

2022 COMP ANNUAL SCIENTIFIC MEETING
B.Côté (1), B.Raaymakers (2), S.Woodings (2), P.Uijtewaal (2), W.de Vries (2), S.Lambert-girard (1), F.Therriault-Proulx (1), M.Fast (2) | Medscint, Canada (1), UMC Utrecht, Netherland (2)

Technical note: Characterization and practical applications of a novel plastic scintillator for online dosimetry for an ultrahigh dose rate (FLASH)

Although FLASH radiation therapy is a promising novel technique, the ultrahigh pulsed dose rates mean that experimental dosimetry is very challenging. The plastic scintillator shows a linear and reproducible response and is able to accurately measure the radiation absorbed dose delivered by 16-MeV electrons at UHPDR. The dose is measured accurately in real time with a greater level of precision than that achieved with a radiochromic film.

Med Phys. 2022
Y.Poirier (1,2), J.Xu (1), S.Mossahebi (1), F.Therriault-Proulx (3), A.Sawant (1) | 1- Department of Radiation Oncology, University of Maryland School of Medicine, Maryland, USA, 2- Department of Medical Physics, McGill University, Quebec, Canada, 3- MEDSCINT, Quebec, Canada

Brachytherapy Technique Commissioning Using the HYPERSCINT Plastic Scintillation Detector

Accurate dosimetry in brachytherapy is not an easy task, as most detectors exhibit volume averaging or energy dependence reducing their usability. Free from these limitations are plastic scintillation detectors, which makes them well suited for brachytherapy applications, either for in vivo dosimetry or commissioning. This work aims to determine if the HYPERSCINT scintillation dosimetry research platform can be used for brachytherapy dose measurement in the context of commissioning a new brachytherapy technique.

2020 AAPM ANNUAL MEETING
M.Goulet, N.Octave, P.Duguay-drouin | CISSS – Chaudiere-Appalaches, Lévis, QC, CA

Beam matching for small-field dosimetry applications using accelerator solenoid current and a miniature plastic scintillation detector.

The goal of this work is to determine the optimal accelerator solenoid current (ASOL) that minimizes the spread of measured small field output factor (OF) values between a series of machines. Small field OF measurements of 6 MV flattening filtered beams from 3 Varian TrueBeam linear accelerators were performed using a 1 mm diameter by 1 mm length plastic scintillation detector HYPERSCINT HS-RP200 research platform. In conclusion Small field dosimetry characteristics are highly sensitive to beam focal spot size. and the possibility to optimize accelerator focusing coil current to reduce OF and penumbra width spread enables new avenues in beam matching of series of machines, especially for SRS and SBRT techniques.

2022 COMP ANNUAL SCIENTIFIC MEETING
L.Gingras (1), F.Beaulieu (1), M.Besnier (1), B.Côté (2), D.Leblanc (2), L.Beaulieu (1), L.Archambault (1) | 1- CHU de Québec – Université Laval, QC, CA, 2- MEDSCINT, QC, CA

Plastic Scintillation Detector for Dosimetric Characterization of Mobetron Ultra-High Dose Rate Electron Beam

The purpose is to present commissioning data for the MOBETRON electron radiation therapy system (IntraOp) at ultra-high dose rate using the HYPERSCINT plastic scintillation detector. The suitability of using a plastic scintillator as an active dosimeter for commissioning measurements of an ultra-high dose rate electron beam has been demonstrated (reference dosimetry, DPP, beam penetration, linearity with number of pulses, linearity with PW and short-term output stability).

2022 AAPM ANNUAL MEETING
G.Famulari (1), K.Zerouali (1), O.Piron (1), JF.Aubry (1), F.DeBlois (1), JF.Carrier (2) | 1- Centre Hospitalier de l’Universite de Montreal (CHUM), Montreal, QC, CA, 2- Departement de Physique, Universite de Montreal, Montreal, QC, CA

Pre-Clinical and Clinical Evaluation of the HYPERSCINT Scintillation Dosimetry Research Platform

The purpose of this work is to evaluate the HYPERSCINT scintillation dosimetry research platform (Medscint Inc., Quebec City, Canada) designed for clinical QA for use in in-vivo dosimetry measurements.

2020 AAPM AM
I.Schoepper, E.Trestrail, S.Dieterich, M.Kent | WR Pritchard Veterinary Medical Teaching Hospital, UC Davis School of Veterinary Medicine UC Davis School Of Veterinary Medicine, Pacific Crest Medical Physics ,Chico, CA, UC Davis Medical Center, Sacramento, CA, Surgical and Radiological Sciences, UC Davis School of Veterinary Medicine, Davis, CA