Investigation of the HyperscintTM Plastic Scintillation Dosimetry System Performance in a 15 MeV electron beam at FLASH dose rates

The performance of the HYPERSCINT plastic scintillation dosimetry system in a 15 MeV electron beam operating at FLASH dose rates was investigated. The linac produced a 15 MeV electron FLASH beam with an average dose rate of ~250 Gy/s and dose per pulse between 1 Gy and 1.7 Gy, at the level just above the multi-leaf collimator. The HYPERSCINT plastic scintillator detector system agreed with OSLDs within 3.5% for 10 to 20 pulse FLASH deliveries using the standard dose rate calibration for both dosimeters.

2021 COMP ANNUAL SCIENTIFIC MEETING
C.Mendez, P.Petric, T.Karan, C.Duzenli | BC Cancer, Vancouver, BC, CA

EFLASH Dosimetry On a Conventional Linac Using Pulse-Gated Delivery

To build on previous experiments and improve reproducibility of electron FLASH delivery on a conventional linear accelerator, a pulse-gating circuit was constructed and tested with several dosimeters including : a 0.01cc volume ion chamber, optically stimulated luminescence dosimeters (OSLDs), Gafchromic MD film and a novel plastic scintillation detector with spectral analysis (HYPERSCINT).

2021 AAPM ANNUAL MEETING
C.Duzenli, C.Mendez, M.Petric, J.Sweeney, D.Ta, T.Karan | BC Cancer, Vancouver, BC, CANADA

FLASH Irradiation of Drosophila Melanogaster Using Low Energy X-Rays

To investigate the capability of low energy x-rays to elicit the FLASH effect, Drosophila melanogaster larvae were irradiated with ultrahigh dose-rate and conventional radiotherapy dose rates using an inexpensive x-ray tube system. Dosimetry was performed with plastic scintillators and radiochromic film, and the differential survival observed in this work suggests that continuous 120 kVp x-rays can induce a FLASH effect.

2022 AAPM ANNUAL MEETING
A.Hart, J.Dudzic, J.Eby, S.Perlman, M.Bazalova-Carter | University of Victoria, Victoria, BC ,CA

Dosimetric Characterization of the ARIEL 10 MV X-Ray Ultrahigh Dose-Rate (FLASH) Irradiation Platform at TRIUMF

The purpose was to characterize the beam delivery capabilities and dose rates achievable on the new ultrahigh dose-rate 10MV x-ray irradiation platform at TRIUMF. Beam commissioning and dosimetry have been conducted on the ARIEL x-ray FLASH irradiation platform using film doses and scintillators. Measured dose rates support that the 10MV x-ray beam may be used as a UHDR source compatible with FLASH radiobiological experiments.

2022 AAPM ANNUAL MEETING
N.Esplen (1), L.Egoriti (2), T.Planche (3), A.Hart (1), B.Paley (3), C.Hoehr (3), A.Gottberg (3), M.Bazalova-Carter (1) | 1- University of Victoria, BC ,CA, 2- University of British Columbia, BC, CA, 3- TRIUMF, BC, CA

Plastic Scintillation Detector for Dosimetric Characterization of Mobetron Ultra-High Dose Rate Electron Beam

The purpose is to present commissioning data for the MOBETRON electron radiation therapy system (IntraOp) at ultra-high dose rate using the HYPERSCINT plastic scintillation detector. The suitability of using a plastic scintillator as an active dosimeter for commissioning measurements of an ultra-high dose rate electron beam has been demonstrated (reference dosimetry, DPP, beam penetration, linearity with number of pulses, linearity with PW and short-term output stability).

2022 AAPM ANNUAL MEETING
G.Famulari (1), K.Zerouali (1), O.Piron (1), JF.Aubry (1), F.DeBlois (1), JF.Carrier (2) | 1- Centre Hospitalier de l’Universite de Montreal (CHUM), Montreal, QC, CA, 2- Departement de Physique, Universite de Montreal, Montreal, QC, CA

Quantifying the DNA-damaging Effects of FLASH Irradiation With Plasmid DNA

The objective is to investigate a plasmid DNA nicking assay approach for isolating and quantifying the DNA-damaging effects of ultrahigh-dose-rate (ie FLASH) irradiation relative to conventional dose-rate irradiation. The doses and dose rates were verified independently using EBT-XD Gafchromic film placed directly above the DNA-based phantom and HYPERSCINT high temporal resolution plastic scintillator placed immediately beside the DNA phantoms (both phantoms had been previously calibrated at conventional dose rates and validated at FLASH-RT dose rates).

INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY, BIOLOGY, PHYSICS
A.Perstin (1), Y.Poirier (2), A.Sawant (2), M.Tambasco (1) | 1- Department of Physics, San Diego State University, CA, USA, 2- Department of Radiation Oncology, University of Maryland School of Medicine, Maryland, USA

Characterization of an x‐ray tube‐based ultrahigh dose‐rate system for in vitro irradiations.

To present an x-ray tube system capable of in vitro ultrahigh dose-rate (UHDR) irradiation of small < 0.3 mm samples and to characterize it by means of a plastic scintillation detector (PSD).

Med Phys. 2021
D.Cecchi (1), F.Therriault-Proulx (2), S.Lambert-girard (2), A.Hart (1), A.Macdonald (1), M.Pfleger (1), M.Lenckowski (1), M.Bazalova-Carter (1) | Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia, Canada, MedScint, QC, CA

Feasibility of Plastic Scintillator Dosimeters for FLASH Therapy

To examine the capabilities of plastic scintillator dosimeters (PSDs) to accurately measure FLASH radiotherapy dose rates delivered with an x-ray tube.

2020 AAPM AM
D.Cecchi (1), F.Therriault-Proulx (2), M.Bazalova-Carter (1) | University of Victoria, Victoria, BC ,CA,, Medscint, QC, CA

Plastic and Lead-Doped Scintillators for Ultrahigh Dose-Rate Irradiations Delivered with An X-Ray Tube

To examine the capabilities of plastic scintillators of different compositions to accurately measure dose in high dose-rate dose irradiations delivered with an x-ray tube.

2021 AAPM AM
D.Cecchi (1), C.Gigeure (2), F.Larose (2), F.Therriault-Proulx (3), L.Beaulieu (2), M.Bazalova-Carter (1) | University of Victoria, Victoria, BC ,CA,, CHU de Quebec – Universite Laval, QC, CA, MedScint, QC, CA

Evaluation of scintillation detectors for ultrahigh dose-rate x-ray beam dosimetry

FLASH-Radiotherapy is an emerging ultrahigh dose rates radiotherapy technique, and animal studies have demonstrated the safety and efficacy of the technique in cancer treatment. A reliable real-time dosimeter system is crucial for the characterization of the so-called ‘FLASH-effect’, and an accurate beam delivery. This study aims to benchmark the performance of optical fiber inorganic scintillating detectors (ISDs) with plastic scintillating detectors (PSDs) for an ultrahigh dose-rate x-ray beam irradiation. Measurements includes : relative scintillator output, signal linearity with dose and dose rate, signal-to-noise ratio (SNR), signal stability and reliability.

The PSDs resulted in the highest reliability for a UHDR beam measurement with a CV of <0.1% while the Gd2O2S:Tb showed excellent repeatability (coefficient of variation (CV) <0.1%) compared to other detectors. All detectors showed good linearity with tube current (R2 < 0.975) and shutter exposure (R2 >0.999).

Proc Spie
Shahirah Shaharuddin (1), Alexander Hart (2), Magdalena Bazalova-Carter (2), Luc Beaulieu (3), Cloe Giguere (3), Christoph Kleefeld (1), Mark J. Foley (1) | 1. National University of Ireland, Galway (Ireland), 2. University of Victoria (Canada), 3 University Laval (Canada)