FLASH-Radiotherapy is an emerging ultrahigh dose rates radiotherapy technique, and animal studies have demonstrated the safety and efficacy of the technique in cancer treatment. A reliable real-time dosimeter system is crucial for the characterization of the so-called ‘FLASH-effect’, and an accurate beam delivery. This study aims to benchmark the performance of optical fiber inorganic scintillating detectors (ISDs) with plastic scintillating detectors (PSDs) for an ultrahigh dose-rate x-ray beam irradiation. Measurements includes : relative scintillator output, signal linearity with dose and dose rate, signal-to-noise ratio (SNR), signal stability and reliability.
The PSDs resulted in the highest reliability for a UHDR beam measurement with a CV of <0.1% while the Gd2O2S:Tb showed excellent repeatability (coefficient of variation (CV) <0.1%) compared to other detectors. All detectors showed good linearity with tube current (R2 < 0.975) and shutter exposure (R2 >0.999).